INST.No.INE-374-P11
- 20 -
・
Measure the temperature of your black body furnace, and program and store the calibrated
temperature of black body furnace in the same way as the programming of calibration data at
zero side. Make sure that the calibration data at span side is greater than the calibration data at
zero side
・
Press [MODE] key once more to get the calibration operation execution screen.
・
Press
key to shift the cursor to “GO”..
・
Press ENT key to start the calibration operation
.
・
When the calibration operation finishes correctly, the message “Completed” appears forseveral
seconds
・
For making the calibrated data effective, restart this thermometer.
7.2 Recovery of calibration data
This function is to recover, after the execution of the user calibration, the calibration data to the previous
data.
This thermometer stores the previous calibration data at the execution of the user calibration.
The following procedure is to recover the calibration data to the previous data stored.
・
Press ON/OFF key while pressing [MODE] key to turn the power supply on.
・
Press MODE key twice to get the zero/span calibration execution screen.
・
Press
key to shift the cursor to “GO”.
・
Press ENT key to get the zero calibration data programming screen.
・
Press MODE key three times to get the calibration data recovery screen.
・
Press [
] key to shift the cursor to “GO”.
・
Press [ENT] key.
After the calibration data recovery finishes correctly, the message “Completed” appears.
・
For making the recovered calibration data effective, restart this thermometer.
When the recovering to the previous calibration data from the current calibration
data is executed, the current calibration data is erased and the recovering operation
can not be executed again until the next user calibration is executed. For initializing
to the default calibration data, refer to [9.5 Initializing to the default condition.].
・
If zero/span calibration data are not stored or if the calibration data at zero side
>
the calibration data at span side, the message “Data
Abnormal” appears
and the calibration operation is not executed.
C A L
.
N
O
O
G
C o m p
d
e
e
t
l
C M P
N
O
O
G
U n
d
o
N
O
O
G
C o m p
d
e
e
l
l
Set up it so in this case, it becomes span calibration data>zero calibration data.
Reference
Remarks
Caution