INST.No.INE-475-P1CE Software Version 1.00
-33-
8.7 Measurement
(1) When ON (enabled) is selected in the sample temperature correction and one calibration curve
number in a group is set, the calibration curve number is automatically switched corresponding
to the sample temperature.
For example, when the sample temperature Ts fulfills the following expression, the calibration
curve number becomes “i”.
T1
≦
…..
≦
Ti
≦
Ts < Ti + 1
≦
…..
≦
T10
(2) However, the measured value Y is not the measured value Y
1
on the calibration curve No. “i”
but becomes the following value interpolated from the measured value Y
i
and the measured
value Y
i+1
on the calibration curve No. “i+1”.
Y = (Y
i+1
- Y
i
) ×
+ Y
i
(3) When the sample temperature Ts is lower than the sample temperature of the first calibration
curve number in the group or when it is higher than the sample temperature of the last
calibration curve number, the following error display appears.
However, the calibration curve, of which sample temperature is set as 999.9, is ignored.
* For the sample temperature lower than the sample temperature of the first calibration curve
number in the group ….. Er15 (Sample temperature too low)
* For the sample temperature higher than the sample temperature of the last calibration curve number
in the group ….. Er14 (Sample temperature too high)
T
s
- T
i
T
i+1
- T
i
8.8.1 Displaying with this detector unit
(1) Set the calibration curve number to 0.
CH 0 ENT
(2) Press key for 2 seconds to display the
sample temperature.
(3) For canceling of the sample temperature
display, press SEL key for 2 seconds or
press CH key for 2 seconds
8.8.2 Displaying with the operator
interface/display unit
(1) Set the calibration curve number to 0.
CH 0 ENT
(2) Display the sample temperature.
MODE 2 0 ENT
(3) For canceling of the sample temperature
display, press SEL key for 2 seconds or
press CH key for 2 seconds
8.8 Displaying sample temperature Ts
8. Sample temperature correction function