7.4 Sample measurement:
For constituent of thickness or coating-thickness
Please measure and pay attention below.
1) Place samples at the same position as in actual measurements, set the calibration curve number to
0, and then read the display (absorbance "x").
2) The visible light is irradiated from this detector unit. Confirm that the samples cover the
measurement area completely.
3) Put the sample tilt by 15
°
from the vertical surface and read the absorbance "x")
4) Measure the thickness and coating- thickness “y” by a micro meter or other measuring methods.
The thickness and coating- thickness “y” in this case is average value measured of 9 points in
measurement area.
Measured value by micro meter
No. Absorbance
"x"
Thickness
“y”
1 2 3 4
5
9
1
2
3
n
I M
S E R I E S
15
°
300mm
Measuring objects
Measurement area
Measured points by micro meter
7. Procedure for creating calibration curves
Caution
Conform the front or rear side to flow directions of the sample.