Triton® Traction
48
operAtion
dts pull pAtterns (iF ApplicAble)
The DTS Pull Patterns, only available with P/N 2842 Triton DTS Card, offer a
selection of various intermittent traction programs. It is up to the clinician
to determine the applicability and appropriate parameters of the presets
as applied to the patient.
Selecting a DTS Pull Pattern
Turn the traction unit off with the On/
Off Switch on the back of the unit.
Remove the Multimedia Card (MMC)
from the front of the traction unit.
NOTE:
The Multimedia (MMC) Card can
be discarded once removed.
Insert the Triton DTS Card into the
Multimedia Card (MMC) Port.
Turn the traction unit on with the On/
Off Switch on the back of the unit.
A message will display on the Touch
Screen User Interface saying,
"Unit has
been changed to a DTS unit. Touch
anywhere to continue."
From the Home Screen, press the DTS
Pull Pattern button.
Choose the desired body area for the
DTS Pull Pattern.
Select one of the five desired DTS Pull
Patterns.
NOTE:
The DTS Pull Pattern parameters
can be edited and saved as the
default Pull Pattern or as a User
Protocol.
(See the User
Protocol Section on Pages
43 - 44 for more information)
Installing DTS Software Upgrade Card