Triton® Traction
53
troubleshooting
error messAges
ERROR
CODE
ERROR
TYPE
DEFINITION
PROBABLE CAUSES
POSSIBLE REMEDY
100
WARNING
Treatment has been running for 8 seconds, but no tension is
detected on the rope.
Too much slack in Traction Cord.
Remove Traction Cord slack.
101
WARNING
Patient pressed Patient Interrupt Switch.
Patient Interrupt Switch button pressed.
Touch the Touch Screen to clear the message.
102
WARNING
Patient Interrupt Switch is unplugged.
Patient Interrupt Switch not properly connected to unit.
Properly connect Patient Interrupt Switch and touch the
Touch Screen to clear message.
104
WARNING
User selected Patient Card button on Utilities Screen, but no
traction treatments were found on the card.
Wrong or bad Patient Data Card inserted into unit.
Insert correct Patient Data Card.
105
WARNING
User selected Save To Card, but no card is inserted into the
unit.
No Patient Data Card is inserted.
Properly insert correct Patient Data Card into unit.
106
WARNING
User selected Save To Card, but the card currently inserted is
not a Patient Card.
Wrong or bad Patient Data Card inserted into unit.
Insert correct Patient Data Card.
107
WARNING
User selected Save To Card, but the card currently inserted
is full.
Memory full on Patient Data Card used.
Save Data to PDMS and erase Patient Data Card.
108
WARNING
User typed in a blank Patient Name.
No Patient ID assigned.
Enter Patient ID.
112
WARNING
Error upgrading Control Board Software.
Unknown
Contact Dealer or DJO, LLC for Service.
113
WARNING
Error upgrading Motor Board Software.
Unknown
Contact Dealer or DJO, LLC for Service.
114
WARNING
User pressed START, but has not yet set the Max Level value
to a value greater than zero.
User has not completed Treatment Setup.
Complete Treatment Setup prior to pressing Start.
115
WARNING
User pressed HOME button from treatment review screen
while a treatment was running.
Home button has been pressed.
Press Stop To End Treatment prior to pressing Home.