Triton® Traction
51
troubleshooting
The Triton traction unit is designed with patient safety in mind. An error can be caused by both external and internal disturbances. Errors can be caused
by disruptions in the power supply (such as a voltage break), and excess or inadequate voltage. Errors may also be caused by patient movement during
the traction therapy.
All traction unit errors and warnings are categorized by three numerical digits: Messages beginning with a 1 are general errors, a 2 symbolizes internal
warnings and errors and a 3 signifies a critical error where the problem has locked up the unit.
(See Warning on Page 54 for more information)
Before calling for service, carefully review this User Manual. In the event you are still unable to correct the problem, contact your DJO, LLC dealer for all
repair service. Be certain to specify your model number, serial number, and a detailed description of the issue you encountered.
problem
possible
remedy
The Power On/Off Switch is in the “ON” position -
• Verify the Mains Power Cord is properly connected to the power supply outlet.
nothing
happens.
• Contact dealer or DJO, LLC for service.
The LED Indicator (Power On/Off) Light glows, but there
• Verify the voltage from the power supply of the electrical outlet is the same as that
is nothing displayed on the LCD, or the LCD displays
is listed on your Voltage Rating and Serial Number Plate.
confusing wording.
The LED Indicator (Power On/Off) blinks -
• The unit is in Screen Saver Mode. Touch the screen or any button to reactivate the
nothing is displayed on the screen.
unit.
The Patient Interrupt Switch is not working.
• Make sure the Patient Interrupt Switch is properly connected to the Patient
Interrupt
Switch
Receptacle.
The Rope Release will not work.
• Release all tension on the Traction Cord.
• If there is more than 2kg (5 lbs or 19N) of tension, the Rope Release button is
disabled.
• If patient is attached, have patient move towards the traction head to release tension.
• Shake the Traction Cord while holding down the Rope Release button.