ASX-1400 Autosampler Operator’s Manual
Using the Autosampler
5–5
Use of mismatched sample vials and sample vial racks may
result in malfunctions or sample spills. Be sure your vials meet
the given requirements.
To order additional supplies, refer to the
CETAC Accessories and
Supplies Catalog
for the ASX-1400 Autosampler.
Arranging the Sample Vial Racks
You can change the arrangement of the sample vial racks to meet your
needs. The ASX-1400 Autosampler accommodates up to four sample
vial racks of 21, 24, 40, 60, or 90 positions each. The autosampler
physically accepts any and all size combinations and arrangements of
sample vial racks. However, you can use racks of mixed sizes only if the
host computer’s software enables you to do so.
For more information about placing sample vial racks in the sample
tray, see Chapter 3, “Installing the Autosampler.”
Incorrectly defining the position count can result in sample spills and
invalid analysis results.
Starting the Autosampler
Once you arrange the sample vial racks and ensure that the
arrangement is correctly defined in the software, you can start the
autosampler and let it run until the sampling sequence is finished. To
do so, complete the following steps:
1)
Ensure that the rinse station is properly connected.
For more information about proper connections, see Chapter 3,
“Installing the Autosampler.”
CAUTION
Summary of Contents for ASX-1400
Page 1: ...ASX 1400 Autosampler Operator s Manual ...
Page 12: ...Contents ...
Page 17: ...Preface ...
Page 23: ...1 Introduction ...
Page 31: ...2 Preparing for Installation ...
Page 35: ...3 Installing the Autosampler ...
Page 52: ...4 Verifying Installation ...
Page 60: ...5 Using the Autosampler ...
Page 68: ...6 1 6 Maintaining the Autosampler ...
Page 76: ...7 Troubleshooting the Autosampler ...
Page 81: ...8 Appendix A ...