ASX-1400 Autosampler Operator’s Manual
Verifying Installation
4–6
8)
Change the bits per second to 9600, set the data bits to eight,
the parity to none, the stop bits to one, the flow control to
none, and click OK.
9)
Select File/Properties.
The COM1_Test Properties box appears.
10)
Select Settings on the COM_Test Properties box and click on
the ASCII setup button on the lower right of the box.
The ASCII setup box appears (Figure 4-5).
Figure 4-5.
ASCII Setup Box.
11)
In the ASCII Setup box, select the following items:
a)
Echo typed characters locally.
b)
Append line feeds to incoming line ends.
c)
Wrap lines that exceed terminal width.
d)
Click OK.