Bedienungsanleitung_CV-EL_DE_EN_REV03.docx
Page 21 of 30
CertoClav Sterilizer GmbH
Peintner Strasse 10
A-4060 Leonding / Austria
Do not transport the device when full.
The device may only be repaired by CertoClav Sterilizer GmbH or by persons authorized by
CertoClav Sterilizer GmbH (specialized company or customer service) using original spare parts
and following the service instructions.
Avoid contact of the aluminum chamber with stainless steel due to chemical processes.
Autoclave – Safety devices
Secure Locking mechanism and steam release valve within the lid
Are providing safety against opening under pressure and will avoid a pressure build up, if the unit is not
closed the correct way. By turning the red knob to “open”, steam will be released. As long as the unit is
not securely locked, the steam release valve remains open and so prevents any build-up of pressure.
Pressure Control valve:
When pressure/temperature is exceeded by about 10% over the highest working pressure (2,7 bar),
steam is released through a safety mechanism. The temperature sinks.
Safety valve:
At 3,6 to 4 bar a safety valve opens in the event the pressure control valve fails. This safety valve closes
automatically after pressure dropped. In any case than the pressure control valve has to be changed, as it
is defect.
Overpressure safety recess
Should all safety devices blocked at an excess pressure at around 4,7 bar the lid gasket would be squeezed
through an overpressure safety recess at the rim of the chamber with a loud noise. If this occurs do not
operate the unit anymore and return it to the dealer or the manufacturer
If this occurs do not operate the unit anymore and return it to the dealer or the manufacturer for
examination and check. It is dangerous to work with such a unit.