39
ORACLE Batch Testing with an Air or Vacuum Oven
5.
Once all the values have been correctly entered, press the NEXT icon to continue.
6.
The total number of batches set in the BATCHING section of the Sample Name page will be listed. Here, each
sample can be edited to provide a unique Sample ID and Tag.
• The Play icon will begin the analysis for the individual sample chosen.
• The Repeat Icon will duplicate the Sample ID and Tag, in the case too few samples were chosen in Step
3 and more are needed. The additional sample will be added to the list directly underneath the Sample ID
from which the duplicate icon was chosen.
• The Delete icon will delete any Sample ID in the case that too many were chosen in Step 3, or any other
reason a sample must be removed from the testing process.
7.
At this time, the user can press the box containing the sample weight, and the system will attempt to register
the sample weight.
• If using the balance, choose Balance, follow the instructions on the screen to Tare the weight of the
sample pad/film, then add the correct amount of sample to the pad/film and record the initial weight.
• If manually inputting the weight, simply type the initial weight that was previously recorded, then press OK
to save the weight.
8.
Once the samples have been fully dried, view “Sample Preparation using the ORACLE Station” section for de-
tailed information on preparing the samples for ORACLE analysis.
9.
Place each sample in the ORACLE tube into the heater block.
10.
Once all the samples have been completely prepared, begin a timer to ensure every sample is heated for the
full optimized conditioning time (typically 1 hour).
11.
Once the samples have been fully heated, press the Play icon next to the corresponding Sample ID.