32
Software Setup
11.
Select the “Tools” header followed by the “System” button.
12.
Once in the System window, choose the Network button, which will bring up a window with two information
boxes.
13.
One box should already be configured for Trac/Coreboard with an IP Address of 10.0.0.1. Do not change the
configuration of this box, use the other box to continue with the remaining steps.
NOTE
If the System Network icon has a red X over it, then the Ethernet cable connecting the SMART 6 and
ORACLE is not correctly plugged in or functioning properly. If the cable is plugged into the correct port and
there is still a red X, please contact CEM Service.
14.
Using the box that does not have an IP address populated, select the gear button to access the pop-up menu,
then choose the SET SMART 6 option to connect the SMART 6. This will set the IP Address connecting the
ORACLE to the SMART 6 unit.