12. TECHNICAL
CONFIGURATION
Press
when the scales are in normal weighing mode and the display will show “F0 H-L”. Press
until
the display shows “PROG” and press
to enter. The display will show “PIN”. Then press
and
to enter Technical Configuration mode and
to select the parameter. Press
to save and
to exit.
FUNCTIÓN
SUB-FUNCTIÓN
DESCRIPTION
AZN 0
This option is used to select zero tracking options.
Options: 0,5d, 1d, 2d, 4d
0-AUTO
This option is used to select the initial zero setting range when
the indicator turns
Options: 0%, 2%, 5%, 10%, 20%
0- RANGE
This option is used to select the manual zero range when the
ZERO key is pressed
Options: 2%, 4%, 10%, 20%, 50%, 100%
P1 REF
Speed
This determines the ADC speed. Press
to select the ADC
speed and
to save it.
7.5: 7.5 times per second
15: 15 times per second
30: 30 times per second
60: 60 times per second
Note: 15 or 30 times per second is recommended
DECI
This option is used to select the decimal point.
Options: 0, 0.0, 0.00, 0.000, 0.0000
INC
This option is used to select the step.
Options: 1, 2, 5, 10, 20, 50
C AP
This display shows xxxxx to set-up capacity. After selecting,
the VC-50 / SVC-50 / MVC-50 / LVC-50 will show the
parameter for animal weighing scales if the external resolution
that has been set is below 1/3000.
First, set the vibration range. Use
to select the desired
value: 0(deactivated)/5/10/15/20/25/30/35/40/45/50, vibrations
in reading data in the range that you determine. The reading
will be set.
Then set-up the reading-block option. A1 (block minimum
data), A2 (block mean data), A3 (block maximum data)
Then set-up the increase value. Press
to enter data,
and then
to save. After setting up this data, when the
readings have been saved, if any items are added to the
platform or any products are removed, the reading will be
updated and saved again.
Then set-up the delay time: 50/60/70/80.
Then set-up the block conditions. 3 may be selected (the VC-
50 / SVC-50 / MVC-50 / LVC-50 will block reading data if it
detects reading data in a vibration range of 3 times) or 8 (it will
block if reading data is detected in a vibration range of 8
times).
P 2 CAL
CAL
Calibration, see details in section 10.
TRI
This display shows xxxxx to trimmer load cells.
COUNT
This display shows
xxxxx to indicate internal counting
.
P3 P RO
RESET
This resets the original factory default set-up.