15.4. TEST REMINDER
This function allows displaying, based on the chosen interval, a message
recalling the performance of the relevant test.
Set whether and when Test reminders are to be activated (Vacuum -
Helix/Bowie&Dick Test- Helix/Bowie&Dick Test) based on the
available options.
Once the fields have been set, activate the reminder with CONFIRM.
Reminders activate at 8 a.m. of the selected day or upon switching-on (if
it occurs after 8 a.m.).
The user can choose between:
- Starting the test;
- Postponing the test (the reminder appears again the next day);
- Ignoring the test (the reminder appears again at the next interval).