78
EN
CYCLE
DESCRIPTION
NOMINAL VALUES
BASIC CYCLE
PARAMETERS
STERILIZABLE MATERIALS
NOTES
Temp
e
ra
tu
re
(
°C)
P
re
s
s
u
re
(
b
a
r)
R
e
te
n
tion
t
im
e
(
mi
n
)
C
y
c
le
t
y
p
e
(
E
N
1
3
0
6
0
:2
0
1
4
)
P
re
-v
a
c
u
u
m
(F=
fr
a
c
tion
a
te
d
;
S
=
s
in
g
le)
S
ta
n
d
a
rd
d
ry
ing
(
mi
n
)
Total
c
y
c
le
t
ime
(Max
f
ill
ing
)
Max
H
2
O
c
o
n
s
u
mp
tion
(
ml
/c
y
c
le)
A
v
e
ra
g
e
e
n
e
rg
y
c
o
n
s
u
m
p
tion
(k
W
h
/c
y
c
le)
TY
P
E
MA
X
TO
TA
L
M
A
S
S
(
k
g
)
MA
X
M
A
S
S
P
E
R
TR
A
Y
(
k
g
)
MA
X
M
A
S
S
P
E
R
I
TE
M
(k
g
)
HELIX/BD
TEST
135
2.2
3.5
-
F
1
24
-
-
Test device only (without
another load)
-
-
-
VACUUM
TEST
-
-0.8
-
-
-
-
18
-
-
Empty chamber
-
-
-
HELIX/BD
TEST
(executable in
sequence)
-
-
-
-
-
-
50
-
-
-
-
-
-
(*) To set a sterilization time of 5.5 minutes, contact the Technical Service.
Single Pre-Vacuum = 1 pre-vacuum; -0.8 bar (see figures in the following pages).
Fractionated Pre-Vacuum = 3 pre-vacuum; -0.8 bar each (see figures in the following pages).
Definition of hollow loads in accordance with standard EN13060:2014.
The term “hollow loads” in this manual refers both to “narrow-lumen” elements (section 3.18 EN 13060:2014) and “simple hollow” elements
(section 3.30 EN 13060:2014).
The term “hollow load B” refers ONLY to “simple hollow” elements (section 3.30 EN 13060:2014).
https://stomshop.pro/