100
EN
CODE
POSSIBLE CAUSE
SUGGESTED SOLUTION
E011
The discharge tank contains water of very poor
quality.
RESET according to the instructions. Empty both tanks IMMEDIATELY,
replace filter elements and fill the filling tank.
E012
The cycle limit after which integrated filters must be
replaced has been reached.
RESET according to the instructions. Empty both tanks IMMEDIATELY,
replace filter elements and fill the filling tank.
E013
The filling tank contains water of very poor quality.
RESET according to the instructions. Empty the filling tank IMMEDIATELY
and replace the demineralisation filter.
In these conditions, the sterilizer allows starting a maximum of 5
consecutive cycles, after which cycles are automatically aborted
until the correct water quality is
detected (<15 μs/cm). This
precaution is necessary to prevent damage to the device.
E020
Door lock mechanism limit microswitch failure.
RESET according to the instructions.
Try restarting the program a second time.
If the problem persists, contact Technical Service (see the Appendix).
Door lock system gearmotor failure.
E021
Door lock mechanism limit microswitch failure.
RESET according to the instructions.
Contact Technical Service (see Appendix).
Door lock system gearmotor failure.
E022
Door lock microswitches failure
RESET according to the instructions.
Contact Technical Service (see Appendix).
E030
Water level in the filling tank below minimum.
RESET according to the instructions.
Top up with water up to the MAX level (or at least up to over the MIN level).
MIN water level sensor failure.
Contact Technical Service
(see Appendix).
E031
Water level in the drain tank over the MAX level.
RESET according to the instructions and empty the tank.
Completely drain the drain tank.
MAX water level sensor failure.
Contact Technical Service
(see Appendix).
Problem in the hydraulic circuit.
E042
Warning that the maximum water level in the tank
has been reached (manual filling)
Interrupt the filling operation to prevent water spillage.
E050
Reminder to run the Vacuum Test cycle
Run Vacuum Test as planned
E051
Reminder to run the Helix Test cycle
Run Helix Test cycle as planned
E052
Reminder to run the combined Helix Test
cycle
Run Helix Test combined cycle as planned
E060
The autoclave cannot connect to Lan network
Make sure that configuration parameters of the Lan network are correct.
Check that the Lan network chosen for the connection is working properly.
Contact Technical Service (see Appendix).
E061
The autoclave cannot connect to Wi-Fi network
Make sure that configuration parameters of the Wi-Fi network are correct.
Check that the router managing the Wi-Fi network is on and that the Wi-Fi
network chosen for the connection is working properly.
Contact Technical Service (see Appendix).
E900
Air seepage through the gasket.
RESET according to the instructions.
Thoroughly clean the gasket with a clean cotton cloth moistened with water.
Restart the program.
Problem in the hydraulic circuit.
Contact Technical Service
(see Appendix).
E901
Excessive humidity in the sterilization chamber.
RESET according to the instructions.
Thoroughly dry the inside of the chamber and restart the program.
Air seepage through the gasket.
RESET according to the instructions.
Thoroughly clean the gasket with a clean cotton cloth moistened with water.
Restart the program.