24.3. HAZARDS (CATEGORY H)
CODE
ALARM DESCRIPTION
MESSAGE ON THE DISPLAY
RESET MODE
H150
MPX pressure sensor broken/not connected
PRESSURE
SENSOR OPEN CIRCUIT
CONTACT TECH.SERVICE
3
H160
MPX pressure sensor short-circuited
PRESSURE
SENSOR SHORT-CIRCUIT
CONTACT TECH.SERVICE
3
H400
Pconv/T ratio not balanced (Pconv>T) (STERILIZATION phase)
INCORRECT P/T RATIO
CHECK LOAD
2
H401
T/Pconv ratio not balanced (T>Pconv) (STERILIZATION phase)
INCORRECT T/P RATIO
CHECK LOAD
2
H402
Temperature over the MAX limit (STERILIZATION phase)
TEMPERATURE BEYOND
MAXIMUM LIMIT
CONTACT TECH. SERVICE
2
H403
Temperature below the MIN limit (STERILIZATION phase)
TEMPERATURE BELOW
MINIMUM LIMIT
CONTACT TECH. SERVICE
2
H404
Floating temperature over the limit (STERILIZATION phase)
ERRATIC TEMPERATURE
CONTACT TECH. SERVICE
2
H405
Pressure over the MAX limit (STERILIZATION phase)
PRESSURE BEYOND
MAXIMUM LIMIT
CONTACT TECH. SERVICE
2
H406
Pressure below MIN limit
(STERILIZATION phase)
PRESSURE BELOW
MINIMUM LIMIT
CONTACT TECH. SERVICE
2
H410
Time measurement error
INTERNAL TIMER ERROR
CONTACT TECH.SERVICE
2
H990
Excessive pressure (sterilization chamber, MPX)
PRESSURE BEYOND
MAXIMUM LIMIT
CONTACT TECH. SERVICE
2
H991
Overheating (sterilization chamber, PT1)
PT1 OVERHEATING
CHECK LOAD
2
H992
Overheating (steam generator, PT2)
PT2 OVERHEATING
CONTACT TECH.SERVICE
2
H993
Overheating (layer resistance, PT3)
PT3 OVERHEATING
CONTACT TECH.SERVICE
2
1 = OK (warning)
2 = OK + door unl RESET
3 = Cycle OK + door unl RESET