![CEFLA MOCOM Supreme-17 Manual Download Page 104](http://html1.mh-extra.com/html/cefla/mocom-supreme-17/mocom-supreme-17_manual_2578944104.webp)
104
EN
CODE
POSSIBLE CAUSE
SUGGESTED SOLUTION
A258
Steam seepage through the gasket.
RESET according to the instructions.
Thoroughly clean the gasket with a clean cotton cloth moistened with
water and restart the program.
Excessive load.
RESET according to the instructions.
Check that the load does not exceed the maximum values permitted.
(See the Summary Table in Appendix Technical Characteristics).
Problem in the hydraulic circuit.
Contact Technical Service (see Appendix).
Steam generator safety thermostat triggered.
Steam generator malfunction.
A260
Drain filter obstructed.
Clean the drain filter (see Maintenance Appendix).
Problem in the hydraulic circuit.
Contact Technical Service (see Appendix).
A261
Bacteriological filter obstructed.
Clean the drain filter (see Maintenance Appendix).
Problem in the hydraulic circuit.
Contact Technical Service (see Appendix).
A262
Bacteriological filter obstructed.
Clean the drain filter (see Maintenance Appendix).
Problem in the hydraulic circuit.
Contact Technical Service (see Appendix).
A353
Drain filter obstructed.
Clean the drain filter (see Maintenance Appendix).
Problem in the hydraulic circuit.
Contact Technical Service (see Appendix).
A356
Drain filter obstructed.
Clean the drain filter (see Maintenance Appendix).
Problem in the hydraulic circuit.
Contact Technical Service (see Appendix).
A360
Drain filter obstructed.
Clean the drain filter (see Maintenance Appendix).
Problem in the hydraulic circuit.
Contact Technical Service (see Appendix).
A362
Drain filter obstructed.
Clean the drain filter (see Maintenance Appendix).
Problem in the hydraulic circuit.
Contact Technical Service (see Appendix).
https://stomshop.pro/