20.2. SCHEDULED MAINTENANCE MESSAGES
The sterilizer periodically displays warning messages relevant to "routine" maintenance operations that must be carried out in order to ensure the proper
operation of the device.
Press OK to confirm that the required maintenance operation has been completed.
Press “REMIND” to postpone the operation.
In this case, the warning message will reappear the next time the sterilizer is used.
Warnings are displayed with the following frequency:
WARNING MESSAGE
BOILER FILTER CLEANING
DOOR LOCK LUBRICATION
DUST FILTER CLEANING
BACTERIOLOGICAL FILTER REPLACEMENT
TANK FILTER CLEANING
BOILER GASKET REPLACEMENT
GENERAL SERVICE
A regular maintenance is essential to achieve the best performance of the device.
Periodically, a message will be displayed requesting that the above maintenance operations are performed.
For further information or in case of doubt, contact the technical service: if they have performed regular maintenance on the device, the technician
might have already carried out some of these operations (e.g. >Replacement of the bacteriological filter or of the gasket).
Always keep in mind the following
general warnings:
•
Do not
wash the sterilizer with direct jets of water, neither under pressure nor sprinkled. Seepage into electrical and electronic components could
damage the functioning of the device or its internal parts, even irreparably;
•
Do not
use abrasive cloths, metal brushes (or other aggressive materials) or products for metal cleaning, both solid and liquid, to clean the device or
the sterilization chamber;
•
Do not
use unsuitable chemical products or disinfectants to clean the sterilization chamber. In fact, these products can cause damages, even irreparably;
•
Do not
allow limescale or residues of other substances to accumulate in the sterilization chamber, on the door and on the gasket, but remove them
periodically. In fact, such residues may damage these parts, besides compromising the operation of the hydraulic circuit components.
The formation of white spots on the base of the internal wall of the chamber means that you are using poor quality demineralised water.
Before performing ordinary maintenance, make sure that the
power cord plug is removed from the mains socket.
If this is not possible, move the external switch of the
device’s power supply line to Off.
If the external switch is distant or not visible to the
mainta
iner, place a “work in progress” sign on the switch,
after turning it off.