Page 60
Electrical Signal Input
Electrical signals at frequencies >2Hz can be applied to the Vexo H instruments
by interfacing a suitable signal generator with an output impedance of 600Ω to
the 5 pin Input Lemo Socket, type (EGG.0B.305.CLL).
Each individual axis shall be subject to the following circuitry, (see Signal Wiring).
AGND
SIGNAL
AGND
SIGNAL
+5V
R1
3K3
C1
1000uF
R2
3K3
SIGNAL
GENERATOR
TRANSDUCER
INPUT SOCKET
Maximum Electrical Signal Input For No Damage
5 Volts (Peak to Peak)
Environmental Stabilization Time
30 minutes
Warm up Time
≤
2 minutes
Settling Time
It is recommended that a calculation settling period of ≥30 seconds is allowed
for in any recording.
Temperature Operating Range
-10
°
C to +50
°
C