Page 57
Chapter 9
Technical Specification
Applicable Standards
ISO 8041:2005 Human Response to Vibration – Measuring Instrumentation
Noise Floors
Axis
Range m/s
2
rms
Low
High
X
0.002
0.02
Y
0.002
0.02
Z
0.002
0.02
Normal Operating Mode
Fitted with Hand Arm accelerometer KD1010
Overload & Under Range Triggering Points
Points when calibrated with an accelerometer of sensitivity 10.0mV/g
Acceleration : Metric m/s
2
RANGE
UR
OL
LOW
0.050
200.0
HIGH
0.500
2000.0
Acceleration : g
RANGE
UR
OL
LOW
0.0051
20.40
HIGH
0.0510
204.0