— 60 —
5-8. Power Consumption Measurement
5-8-1. Test equipment con
fi
guration
5-8-2. Measurement items and speci
fi
cations
Item
Supply voltage
Current value
Run mode
4.0V
200 mA or lower
Sleep mode
4.0V
65 mA or lower
Main Dummy load
4.0V
70-100 mA
OFF
4.0V
1 mA or lower
Laser Scanner movement
4.0V
150 mA or lower
Linear Imager movement
4.0V
150 mA or lower
2D Module movement
4.0V
360 mA or lower
5-8-3. Run mode
No
Operation
Display
Inspection contents
1
Set the supply voltage to
4.0V.
Select [Run mode
current] in the test
program.
Backlight is turned off.
Check that the current value is as follows.
- “200 mA” or lower
Within spec: Press the “1” key.
Out of spec: Press the “0” key.
Pressing “1” or “0” returns you to the
Menu screen.
DT-X100
DC Ammeter
Direct current stabilized power source
Battery terminal
Tape down the battery
cover sensor button.