27 |
P a g e
cascadesciences.com
|
Tel. 503 847-9047
INSTALLATION
C
ONNECT TO A
G
AS
B
ACKFILL
S
UPPLY
Optional:
A clean or inert gas supply—such as nitrogen (N
2
) or argon—may be connected to the vent
intake port located on the back of the oven. This supply will backfill the evacuated oven chamber
when the vent intake valve is opened on the front control panel.
The maximum allowed gas backfill pressure is 15 psi.
Intake Vent ½” OD – To the
oven chamber
CVO-10
Exceeding 15 psi of gas backfill
pressure may damage the oven.
Intake Vent ¼” OD – To the
oven chamber
CVO-5s
Gas Supply
Gas Supply
Intake Vent
Vent