USER MAINTENANCE
End of procedure
Calibration continued
7
Wait 30 minutes for the oven to stabilize
after the
oven has achieved the setpoint
with the corrected
display adjustment.
•
Failure to wait will result in an inaccurate oven
display reading.
8
Once the temperature has stabilized, compare the reference
device and the oven display temperature reading again.
a.
If the readings are the same, or the difference between
the two falls within the acceptable range of your protocol,
the oven is calibrated for temperature.
-Or-
b.
Advance to step 9.
Reference Device
9
If the two readings still fall outside the acceptable range of your
protocol, repeat steps 3 – 8 up to two more times.
•
Three calibrations attempts may be required to
successfully calibrate ovens more than ± 3
⁰
F out of
calibration.
Reference Device
If the temperature readings of the oven and the reference device fall outside your protocol after
three calibration attempts, contact
or your distributor for assistance.
Summary of Contents for CVO-10
Page 1: ...Vacuum Purge Ovens Installation Operation Manual CVO 5 CVO 5 2 CVO 10 ...
Page 4: ...4 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS 53 ...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS ...
Page 10: ...10 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS ...
Page 18: ...18 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN ...
Page 28: ...28 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION ...
Page 30: ...30 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS ...
Page 52: ...52 P a g e cascadesciences com Tel 503 847 9047 USER MAINTENANCE ...
Page 55: ...55 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS ...
Page 56: ......