INSTALLATION
R
EQUIRED
A
MBIENT
C
ONDITIONS
These units are built for use indoors at room temperatures between
15
°
C and 40
°
C (59
°
F and 104
°
F)
, at
no greater than
80% Relative Humidity
(at 25
°
C / 77
°
F). Operating outside these conditions may
adversely affect the unit temperature performance.
When selecting a location to install the unit, consider all environmental conditions that can impact its
temperature performance. These include:
•
Proximity to other ovens, autoclaves, and any device that produces significant radiant heat
•
Heating and cooling vents or other sources of fast-moving air currents
•
High-traffic areas
•
Direct sunlight
R
EQUIRED
C
LEARANCES
All Ovens
These clearances are required to provide air flows for ventilation and cooling.
6 inches (152 mm)
of clearance is required on the sides.
12 inches (305 mm)
of headspace clearance is required between the top of the unit and any
overhead partitions.
Do not place objects on top of the oven.
A KF-25 vacuum port is located on the back of the oven for introducing vacuum-rated thermocouple
feedthroughs into the chamber or connecting to an external vacuum supply source. Leave sufficient
clearance for operators to safely access this port.
6” (152mm)
Door Swing
12” (305mm)
12” (305mm)
6” (152mm)
12” (305mm)
Vacuum Pump
Power Outlet
Summary of Contents for CVO-10
Page 1: ...Vacuum Purge Ovens Installation Operation Manual CVO 5 CVO 5 2 CVO 10 ...
Page 4: ...4 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS 53 ...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS ...
Page 10: ...10 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS ...
Page 18: ...18 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN ...
Page 28: ...28 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION ...
Page 30: ...30 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS ...
Page 52: ...52 P a g e cascadesciences com Tel 503 847 9047 USER MAINTENANCE ...
Page 55: ...55 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS ...
Page 56: ......