61
(2) CAL mode
CH 11
▣
Reason
The resolution exceeds the 1/10,000.
☞
Troubleshooting
Lower the resolution. Change the maximum capacity in CAL1 or change the division in
CAL2 so that the resolution should be below 1/10,000.
CH 12
▣
Reason
The weight for span calibration is lower than 10%, or greater than 100% of the
maximum capacity of the scale.
☞
Troubleshooting
The weight for span calibration should be within 10%~100% of the maximum capacity
of the scale in CAL3.
CH 13
▣
Reason
Load cell output is too small or large at span calibration.
☞
Troubleshooting
Calibrate with lower resolution.
CH 14
▣
Reason
Load cell output is too small or large at zero calibration.
☞
Troubleshooting
Check whether the platform is empty.
Calibrate again after checking in A/D TEST mode.