27
CAL 5
FUNCTION : Span Calibration
KEY
VF
DISPLAY
SUB
DISPLAY
DESCRIPTION
: span calibration
: exit
LOAd
- - - -
SUCCESS
CAL5 LoAd
Load the weight which was set
in CAL 3 and press SET key.
Under span calibration...
Span calibration is completed.
Check whether the displayed
weight is same with setting
weight.
Note 1. If Span calibration is done without any error, SUCCESS message is displayed.
The weight of setting weight is displayed on VFD screen. Check the weight.
Note 2. If the span is low, Error message (CH 13) is displayed. In that case, Calibrate with
Lower resolution. Please check the span value to be resolution × 20 in TEST 3.
CAL 6
FUNCTION : Check if the calibration is done properly
KEY
VF DISPLAY
SUB DISPLAY
DESCRIPTION
: save and
go to next menu
: exit from the
CAL mode
5000kg
▽▽▽▼▽▽▽
CAL6 VErIFY
5000 kg
Note 1. If the central lamp lights up as above VF Display, the bias is "0", and each lamp
indicates the bias of -3, -2, -1, 0, 1, 2, 3 from the left lamp.
Note 2. Confirm if the displayed weight is equal to the setting weight you entered in CAL3,
and press the
key 2 times to go to weighing mode.