28
CAL 7
FUNCTION : Input weight constant calibration after selecting national code
KEY
VF
DISPLAY
SUB DISPLAY
DESCRIPTION
~
:
enter password
&
:
finish the CAL mode
PASS
CAL7 FACtor
Type the password.
Note 1. National code is n = 0 Standard, n = 1 Tailand, n = 2 Turkey
Note 2. Users do not have to use this menu, since it is used for calibration test without a
weight.
Note 3. If you press the
key two times, the display shows “CAL END” and then you
can go to weighing mode.