,175$0$;3/86
*%
&DUORGH*LRUJL
With reference to Fig. 2, connect the footswitch to socket (4) on the rear
panel by inserting the plug and twisting the locking ring;
Insert the saline solution stand into upper socket (8) and then hang the
saline solution bottle by the plastic cradle.
)LJ
With reference to Fig. 3 open the side door of peristaltic pump (1) and
remove the tube (keeping it safe for future uses with irrigators other than
the one supplied); insert the larger section tube (2) (included with the
disposable irrigator supplied) into the peristaltic pump, paying attention to:
- the positioning of supports (3), since, as can be observed in the figure,
the supports perform a restraining function on the exterior of the
pump;
-
the connection of the "IN" and "OUT" sides (4) respectively to the
saline solution bottle and the surgical handpiece (not supplied).
)LJ