Once you have selected a program to edit, move around the fields using
the "shift" key (7 in Fig. 6) and change the relative parameter values with
the "
n
" (5 in Fig. 6) and "
p
"
(6 in Fig. 6)
keys; when you have finished the edit
procedure to save the program return to the "program n°" field and then
quit the program by going to the next or the previous program by pressing
"
n
" (5 in Fig. 6) and "
p
"
(6 in Fig. 6). If the appliance is switched off before
this final step the changes to the parameter values will not be saved in the
memory and the new program settings will be lost.
To edit a program using the footswitch (see Fig. 7 and 8), press the
yellow button (9 in Fig. 7), bring the flashing cursor on the “Step” field (1 in
Fig. 8) by means of lever (10 in Fig. 7) and select the first free or unused
program using the green (3 in Fig. 7) and red (4 in Fig. 7) buttons.
Once you have found the program to be edited, move around the field
using lever (10 in Fig. 7), editing the parameter values using the green (3
in Fig. 7) and red (4 in Fig. 7) buttons; when the setting is complete, to
save the program, return to the “Step” field (1 in Fig. 8) and quit the
program, by moving to another program by pressing the green (3 in Fig.
7) and red (4 in Fig. 7) buttons. If the appliance is switched off before this
final step the changes to the parameter values will not be saved in the
memory and the new program settings will be lost.
A maximum of 32 programs can be saved in memory.
1RWH
The parameter on which the cursor is positioned continues to flash.
ALWAYS SWITCH OFF THE APPLIANCE WHEN YOU FINISH WORKING WITH
IT AND FOLLOWING ALL STOPPAGES, IRRESPECTIVE OF THE MOTIVE.
IF YOU NOTICE MALFUNCTIONS, STOP THE APPLIANCE IMMEDIATELY AND
DISCONNECT IT FROM THE ELECTRICAL POWER SUPPLY. BEFORE
RECONNECTING THE POWER SUPPLY FIND THE CAUSE OF THE FAULT AND
REMEDY IT. IN THE CASE OF PROBLEMS, CONSULT CARLO DE GIORGI.
0$,17(1$1&(
Periodically check the condition and operation of the accessories and parts
subject to wear and tear; in particular check the condition of the burs and the
micromotors in order to identify any possible damage.
Periodically check the condition of cable insulation. The appliance cables
must be replaced exclusively by suitably skilled and authorised personnel.