PREDICTED VALUE SET UP
To change the Predicted Value Settings, from the default shown on the
box, hold down the On
and Enter
keys together, and only
release when ‘Prd’ is displayed. Use the Scrolling Up
/ Down
keys to choose an alternative Predicted Value set.
1 ECCS
4
2 NHANES III
5
3 Asian (Chinese)
6
When the desired number is selected press the Enter
key to
accept and save that predicted value set as default.
VERSION
To check the version number, hold down the On
and Scrolling
Down
keys together, and only release when ‘rE’ is displayed. The
version number will then be shown.
TROUBLE SHOOTING INFORMATION
Should you encounter problems operating your Pulmolife unit, please
consult the table below:
Problem
Possible
Cause
Solution
Does not register a blow
Turbine
incorrectly
fitted
Ensure turbine is tightly
pushed into the unit.
Unit does not turn on or
Solid “bAt” indicator is
displayed, while unit
beeps three times and
powers itself down after
5 seconds
Battery is
exhausted
Replace battery
Calibration Check failed
or cannot be completed
There are
leaks in the
syringe or
connections
Check the syringe and
connections for leaks
Turbine
may be
faulty
Repeat calibration
procedure, if problem still
persists, replace turbine
Shaft of the
syringe
being
pushed
down
The syringe should be
emptied and filled with one
smooth stroke, avoid
pushing down on the shaft
or banging at the end of
each manoeuver.
Summary of Contents for PulmoLife
Page 2: ......
Page 4: ......
Page 44: ...40 Deutsch 5 Sekunden selbstst ndig ab...
Page 119: ...115 VII PULMOLIFE PulmoLife 1 1 1 1 1 2 PULMOLIFE...
Page 120: ...116 WEEE IPX0 PulmoLife 35...
Page 121: ...117...
Page 122: ...118 ECCS NHANES III 45 ECCS NHANES III 0 00 PulmoLife 0 00 1 1...
Page 123: ...119 3 1 80 1 20 90 30 90 1 100 3 CareFusion PeraSafe 36 SSC5000A PulmoLife 36 SST1000...
Page 124: ...120 PeraSafe 10 PulmoLife PulmoLife 3 36 SM2125 Ch 0 00 3 3 3 2 91 3 09...
Page 125: ...121...
Page 126: ...122 MicroGard PulmoLife 3 bAt 5 3...
Page 127: ...123 Prd 1 ECCS4 2 NHANES III5 3 6 rE PulmoLife bAt 5...
Page 128: ...124 IEC 60601 1 B 3 CR 2450 IEC 60601 1 B...
Page 137: ...133 0086 93 42 WEEE SGS Testing Services OSHA...
Page 138: ...134 Enter IX PULMOLIFE PulmoLife FEV1 FEV1 FEV1 FEV1 COPD1 2 PULMOLIFE...
Page 139: ...135 WEEE IPX0 PulmoLife COPD COPD 35 COPD...
Page 140: ...136 Enter ECCS NHANES III 45 ECCS NHANES III 0 00 PulmoLife 0 00 FEV1 FEV1...
Page 142: ...138 PulmoLife Pulmolife 3 36 SM2125 Ch 0 00 3 3 3 2 91 3 09 MicroGard...
Page 143: ...139 PulmoLife 3V bAt 5 3V EU...
Page 144: ...140 Enter Prd 1 ECCS4 2 NHANES III5 3 6 Enter rE Pulmolife bAt 5...
Page 145: ...141 IEC 60601 1 B 3V CR 2450 IEC 60601 1 B...
Page 153: ...149 B 0086 93 42 EEC WEEE Rx SGS...
Page 155: ......