8200-M628-O1 Rev A
212G/212GX Installation & Technical
Page 41
CALIBRATION “C” NUMBERS
The “C” numbers are displayed only during the Test mode operation by pressing the
*
key and
then the
UNITS/TEST
key. The "C" numbers are shown at the end of the test operation and
each number is displayed for approximately 4 seconds, allowing you to record them. Each
number may be up to three (3) digits in length. By recording these numbers you will be able to
return the indicator to its present calibration settings without using test weights simply by
entering the “C” numbers.
Refer to the Setup and Calibration, "
CAL
" section of this manual for
instructions on using the "C" numbers
.
If any components have been changed that affect calibration and/or your scale
is used in a commercial application and must be "Legal for Trade" you can not
use "C" numbers to re-calibrate.
ACCUMULATORS
To
view
the NET accumulator:
1. Press the
*
key and then the
NET/GROSS
key.
2. Press the
*
key to return to normal operation.
To
the NET accumulator:
1. Press the
*
key, the
NET/GROSS
key and then the
key
2. The indicator will return to normal operation when printing has been completed.
To
clear
(zero) the NET accumulator:
1. Press the
*
key, the
NET/GROSS
key and then the
ZERO
key
2. Press the
*
key to return to normal operation.
To
view
the Gross accumulator:
1. Press the
*
key and then the
NET/GROSS
key twice.
2. Press the
*
key to return to normal operation.
To
the Gross accumulator:
1. Press the
*
key, the
NET/GROSS
key twice and then the
key
2. The indicator will return to normal operation when printing has been completed.
To
clear
(zero) the Gross accumulator:
1. Press the
*
key, the
NET/GROSS
key twice and then the
ZERO
key
2. Press the
*
key to return to normal operation.
To
view
the Count accumulator:
1. Press the
*
key and then the
COUNT/SAMPLE
key.
2. Press the
*
key to return to normal operation.
To
the Count accumulator:
1. Press the
*
key, the
COUNT/SAMPLE
key and then the
key
2. The indicator will return to normal operation when printing has been completed.
To
clear
(zero) the Count accumulator:
1. Press the
*
key, the
COUNT/SAMPLE
key and then the
ZERO
key
2. Press the
*
key to return to normal operation.
Summary of Contents for 212G
Page 2: ...8200 M628 O1 Rev A 212G 212GX Installation Technical Page 2 ...
Page 10: ...8200 M628 O1 Rev A 212G 212GX Installation Technical Page 6 INSTALLATION CONT Figure No 2 ...
Page 27: ...8200 M628 O1 Rev A 212G 212GX Installation Technical Page 23 Lamp test on power up ...
Page 52: ...8200 M628 O1 Rev A 212G 212GX Installation Technical Page 48 ...