8200-M628-O1 Rev A
212G/212GX Installation & Technical
Page 28
SETUP AND CALIBRATION, CONT.
A-d
(
A-d?)
- Analog to Digital Filtering
SErSCL
–
(REMOTE INDICATOR SETUP)
If the indicator is to function as a remote indicator, press the
YES
key at the
SErSCL
prompt. If not then press the
NO
key at the prompt.
If
SErSCL
was set to
YES
then, an additional prompt has also been added in the
SIO
section for selecting the port for the LOCAL/REMOTE communications.
At the
LrP
prompt enter either
1
for port 1 or
3
if the optional 2xx-RS232 card is
being used.
IMPORTANT!
If a 200, 205, 210 or 215 indicator is to be used as the LOCAL and as the REMOTE,
then both indicators
MUST
have Rev 2.6.3 or greater software installed.
If a combination of 200, 205, 210 or 215 and a 225 indicator are to be used in a
LOCAL/REMOTE configuration, then the remote 200, 205, 210 or 215 indicator
MUST
have Rev 2.6.1 or greater software installed, and the 225 indicator
MUST
have Rev
1.0.K or greater software installed.
dFLt=
(Digital Filtering)
With
A-d
(
A-d?)
displayed, press the
ENTER
key. The display will change to
dFLt=
. Press the
ENTER
key to show the current value. If the setting displayed is acceptable, press the
ENTER
key to save it. Otherwise, using the numeric keys enter the new setting and then press the
ENTER
key to save it. Allowable values are: 0, 1, 2 or 3. Note that if you select 3 (Custom
Filtering) two additional prompts will be displayed.
dFLt=
0
Disabled – NO Filtering
1
MINIMAL FILTERING (sample rate = 2)
2
MODERATE FILTERING (sample rate = 1)
3 CUSTOM
FILTERING
NOTE!
The prompts,
F=
(Filter Level) and
b=
(Break Range) will only be displayed if you
selected 3 (Custom Filtering) for the
dFLt=
(Digital Filtering) prompt.
F=
(Filter Level)
Press the
ENTER
key to show the current setting for the filter level. The filter level is a number
from 1 to 99 that corresponds to the level of filtering with 1 the least and 99 being the greatest
filtering. To accept the value displayed, press the
ENTER
key, otherwise, use the numeric
keys to enter a new value and then press the
ENTER
key to save it.
b=
(Break Range)
Press the
ENTER
key to show the current setting for the break range. The break range is a
number from 1 to 255 that corresponds to the number of division change to break out of the
filtering. Press the
ENTER
key to keep the displayed value or use the numeric keys to enter a
new value and then press the
ENTER
key to save the new setting. Note that entry of a 0
disables this feature.
Sr=
(Sample Rate)
Press the
ENTER
key to show the current setting for the sample rate. The value displayed is
the sample rate in samples per second. Press the
ENTER
key to save the displayed value or
use the numeric keys to enter a new value (1 to 100) and then press the
ENTER
key to save it.
Summary of Contents for 212G
Page 2: ...8200 M628 O1 Rev A 212G 212GX Installation Technical Page 2 ...
Page 10: ...8200 M628 O1 Rev A 212G 212GX Installation Technical Page 6 INSTALLATION CONT Figure No 2 ...
Page 27: ...8200 M628 O1 Rev A 212G 212GX Installation Technical Page 23 Lamp test on power up ...
Page 52: ...8200 M628 O1 Rev A 212G 212GX Installation Technical Page 48 ...