Star Biphasic Waveform
Version 112-2025-109 B
7-10
* All values are typical.
**Allowable energy range.
Table 16: High Variable Energy (300 VE) Powerheart G3 Waveform
Phase 1
Phase 2
Patient’s
Impedance
(Ohms)
Voltage*
(Volts)
Duration*
(MS)
Voltage*
(Volts)
Duration*
(MS)
Energy**
(Joules)
25
1869
3.3
997
3.2
260-351
50
1906
4.5
1220
3.2
230-311
75
1918
5.8
1306
3.2
210-283
100
1925
7.0
1351
3.2
195-263
125
1928
8.3
1378
3.2
184-248
150
1931
9.5
1396
3.2
176-238
175
1933
10.8
1408
3.2
170-230
Summary of Contents for Powerheart AED G3 Plus 9390A