Scheduled Maintenance
Version 112-2025-109 B
6-4
3
Close the lid.
4
Confirm that the STATUS INDICATOR turns RED.
5
Open the lid and confirm that the Pad indicator is lit.
6
Reconnect the pads and close the lid.
7
Make sure the expiration date is visible through the clear window of the
lid.
8
Check to make sure that the STATUS INDICATOR is GREEN. If the
pads are not installed properly, the PAD indicator will illuminate; call
Customer Service for assistance.
9
Open the lid and confirm that no diagnostic indicators are lit.
10
Check the expiration date of the pads; if expired, replace them.
11
Check the pads packaging integrity.
12
Close the lid
Check the Integrity of the Service Indicator (LED) and Circuitry:
1
Immediately after opening the AED lid, press and hold the Shock button
and confirm that the Service LED is lit.
2
Release the Shock/Continue button.
3
Close the lid.
4
Verify that the STATUS INDICATOR remains RED.
5
Open the lid and confirm that no diagnostic indicators are lit.
6
Close the lid.
7
Verify that the STATUS INDICATOR turns GREEN.
Check the Integrity of the Case:
Examine the molded case of the AED for any visible signs of stress. If the
case shows signs of stress, contact Cardiac Science Customer Service
(See Contact Information on page 1-1) or contact your local Cardiac Science
distributor.
!
CAUTION: Case Cleaning Solutions.
When disinfecting the case, use a non-oxidizing disinfectant, such as ammonium
salts or a glutaraldehyde based cleaning solution, to avoid damage to the metal
connectors.
Summary of Contents for Powerheart AED G3 Plus 9390A
Page 1: ...OPERATOR AND SERVICE MANUAL Version 112 2025 109 B Revised February 4 2010 www AmericanAED com...
Page 6: ...Table of Contents Version 112 2025 109 B vi...
Page 10: ...112 2025 109 B Limited Warranty x...
Page 60: ...Step 6 Post Rescue Version 112 2025 109 B 4 8...
Page 70: ...Frequently Asked Questions Version 112 2025 109 B 6 8...