Parameters
Version 112-2025-109 B
7-2
Environmental Operation
and Standby Conditions
Temperature: 0°C to 50°C (32°F to 122°F)
Humidity: 5% to 95% (non-condensing)
Pressure: 57kPa (+15,000ft) to 103kPa (-500ft)
Shipment and Transport
environmental Conditions
(for up to 1 week)
Temperature: -30°C to 65°C (-22°F to 149°F)
Humidity: 5% to 95% (non-condensing)
Pressure: 57kPa (+15,000ft) to 103kPa (-500ft)
Pads
Self-adhesive, disposable defibrillation pads
Minimum combined surface area: 228cm
2
Extended length of lead wire: 1.3m
9146 Lithium Battery
Specifications
(p/n: 9146-001)
Output voltage: 12VDC (max)
Batteries are non-rechargeable
Lithium contents: 9.2g (max)
Check local regulations for disposal information
Full Operational Replacement Guarantee (from
date of installation): 4 Years
Estimated Shelf Life (from date of manufacture): 5
Years
Typical Shocks: up to 290 shocks
Note:
The battery operating life depends on the
type of battery, actual usage and environmental
factors.
Batteries and Capacitor
Charge Times
A new battery typically takes 10 seconds to
charge the AED to maximum energy.
A battery with reduced capacity causes the red
LED light to initially turn ON and typically takes
13 seconds to charge a fully discharged AED to
maximum energy.
Table 13: Parameters (continued)
Parameter
Detail
Summary of Contents for Powerheart AED G3 Plus 9390A
Page 1: ...OPERATOR AND SERVICE MANUAL Version 112 2025 109 B Revised February 4 2010 www AmericanAED com...
Page 6: ...Table of Contents Version 112 2025 109 B vi...
Page 10: ...112 2025 109 B Limited Warranty x...
Page 60: ...Step 6 Post Rescue Version 112 2025 109 B 4 8...
Page 70: ...Frequently Asked Questions Version 112 2025 109 B 6 8...