Version 112-2025-109 B
Technical Data
7-7
Cardiac Rhythms Used to
Test the Rhythm
Recognition Detection
System for Powerheart G3
AEDs
Shockable Rhythm* – VF: Meets AAMI DF 39
requirement and AHA recommendation of
Sensitivity of >90%
(*Automatic External Defibrillators for Public
Access Defibrillation: Recommendations for
Specifying and Reporting Arrhythmia Analysis
Algorithm Performance, Incorporating New
Waveforms and Enhancing Safety, AHA AED Task
Force and approved by the AHA Science Advisory
and Coordinating Committee. Circulation,
1997(95), pp 1677-1682)
Shockable Rhythm – VT: Meets AAMI DF 39
requirement and AHA recommendation of
Sensitivity of >75%
Non-shockable Rhythm – NSRMeets AAMI DF 39
requirement (>95%) and AHA recommendation
(>99%) of Specificity
Non-shockable – Asystole: Meets AAMI DF 39
requirement and AHA recommendation of
Specificity of >95%
Non-shockable: Meets AAMI DF 39 requirement
and AHA recommendation of Specificity – all
other rhythms of >95%
For detailed information contact Cardiac Science
for white papers:
•
P/N 112-2013-001 (Pediatric Defibrillation
Instructions)
•
P/N 110-0033-001 (RHYTHMx White Paper)
•
P/N 400781 (STAR Biphasic White Paper)
Table 13: Parameters (continued)
Parameter
Detail
Summary of Contents for Powerheart AED G3 Plus 9390A
Page 1: ...OPERATOR AND SERVICE MANUAL Version 112 2025 109 B Revised February 4 2010 www AmericanAED com...
Page 6: ...Table of Contents Version 112 2025 109 B vi...
Page 10: ...112 2025 109 B Limited Warranty x...
Page 60: ...Step 6 Post Rescue Version 112 2025 109 B 4 8...
Page 70: ...Frequently Asked Questions Version 112 2025 109 B 6 8...