Version 112-2025-109 B
Technical Data
7-3
AED Self test Sequence
Daily: Battery, pads, internal electronics, Shock/
Continue or button, and software (no charge).
Weekly: Battery, pads, internal electronics, Shock/
Continue, and software (partial charge).
Monthly (every 28 days): Battery under load,
pads, internal electronics, full-energy charge
cycle, Shock/Continue button, and software (full
charge).
Open Lid (when lid is opened): Battery, pads,
internal electronics, Shock/Continue button, and
software.
Close Lid (when lid is closed): Battery, pads,
internal electronics, Shock/Continue button, and
software.
Table 13: Parameters (continued)
Parameter
Detail