Note: At least one measurement protocol must be set up before any patient can be added
to the Patient Directory or any other function can be accessed from Figure 7-1 Main
Thyroid Uptake Screen.
QUICK STEP GUIDE FOR THYROID UPTAKE PROCEDURE
1. Set up a protocol for measurement (refer to
on page 7-5).
2. Add all the required patient information and dosage information (refer to
on page 7-10).
3. If the counting method selected in protocol is Decay Correct Administered Dose (refer to
on page 7-17), count each capsule for an initial capsule
count before the patient arrives.
4. If needed, measure pre-dose patient when the patient arrives.
5. Administer the capsule(s) to the patient. Record the administered time (refer to
on page 7-4).
6. Count the patient after a desired uptake time (refer to
page 7-19).
7. Print the patient report if needed (refer to
on page 7-24).
MEASUREMENT PROTOCOLS
Protocols need to be set up to select the method of measurement and dose administration to
the patient (refer to
on page 7-5).
Counting Methods
There are two counting methods that can be used as the measurement protocol:
1. Decay Correct Administered Dose
In this method, a single capsule is used.
•
Measure the capsule
before
administering it to the patient.
•
Then administer this capsule to the patient.
•
Measure the patient for uptakes at the required intervals.
Sometimes, individual capsules may vary in activity from one to another. This method
eliminates any error caused by a difference in activity between the capsule
administered to the patient and a second different capsule used as a standard. It is
less time consuming since it eliminates the need to recount a standard capsule each
time an uptake measurement is performed. This method is also cost effective since it
eliminates the need of a second capsule to be used only as a standard. If your facility
uses an alternate method, try making comparison measurements with your current
November 13
THYROID UPTAKE
7-3
Summary of Contents for CAPTUS 3000
Page 2: ......
Page 14: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 22: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 86: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 124: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 266: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 284: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 292: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 304: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 334: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 336: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 337: ......
Page 338: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...