CAPINTEC, INC.
CAPTUS
®
3000
2. To edit values, click in the desired text box and then type in the new number. To change
the Report Font Size, click the desired radio button.
3. Click the
button or press
Alt+D
to save changes or click the
button
or press
Alt+C
to cancel the changes and return to Figure 7-1 Main Thyroid Uptake
Note: Except for Normal Range values, the changed setup is applicable to only new
patients added after the change.
QUICK TIPS:
•
To restore the default values for ROI at any time, click the
button.
•
To count for a time less than the
Counting Time
entered in the setup, you may
stop and save the data midway while measuring.
MAKING MEASUREMENTS
•
Before making any measurements on a patient, verify that all required patient information
has been entered.
•
Any changes to the measurement setup/protocols must be done
before
performing the
uptake.
1. To select a patient for the uptake procedure, click on that record in the
Patient Directory
box on Figure 7-1 Main Thyroid Uptake Screen. The
Patient Status
box will show the
current status on the patient. If this box reads, “All Data has been Entered”, the uptake
procedure can be started. If an uptake procedure is midway for a patient, the box will
display the results obtained to that point.
Note: To select a patient, use the
keys or Page Up and Page Down keys to move
up or down in the patient list. If the list is long, the scroll bar is also available. To
reach a particular patient, press the first letter of the patient name. Press the
key if necessary, to go further down.
If the Patient Status box displays “Data Partially Entered”, no measurements can
be made until all the required information is entered. Refer to Editing Patient
Information on page 7-13.
2. To begin the measurements, follow the sequence of steps as shown below depending
upon the Counting Method selected in the protocol:
a. Decay Correct Administered Dose
i. Click the
button on Figure 7-1 Main Thyroid Uptake
Screen. Follow the on-screen instructions to:
•
If
Pre-Dose Measurement
is selected in the protocol, then you will be
stepped through the measurement of patient background and neck.
November 13
THYROID UPTAKE
7-17
Summary of Contents for CAPTUS 3000
Page 2: ......
Page 14: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 22: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 86: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 124: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 266: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 284: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 292: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 304: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 334: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 336: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 337: ......
Page 338: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...