CAPINTEC, INC.
CAPTUS
®
3000
Figure 10-46
Default Values
frame
Administered to Standard Activity Ratio
– This is the standard dilution factor,
which is 100 for this procedure.
Administered Volume
– This is the volume injected into the patient.
Counting Time
– This is the duration for counting, in seconds, each time a
measurement is performed. The default value is set to 60 seconds.
Repetitions
frame – This is the number of repetitions each time a measurement is made.
For two repetitions, the average value is used in calculations. The second repetition may
be skipped during the measurement procedure, if so desired.
Normal Range (Female / Male)
frames –
Minimum
and
Maximum
(ml/kg)
for
Whole
Blood, Plasma
and
Red Blood Cells
– These numbers will define a range of values for
blood, plasma and red blood cell volumes respectively, for Females and Males, that can
be considered as medically normal for the patient’s weight. They are entered as seen fit
by the doctors. Out of range results indicate a medical condition and will be flagged in the
report. Entering these values is optional.
2. To input values in any frame, click in the appropriate text box or radio button or use the
Tab
key to move from one selection to another.
10-32
LAB TESTS
November 13
Summary of Contents for CAPTUS 3000
Page 2: ......
Page 14: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 22: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 86: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 124: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 266: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 284: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 292: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 304: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 334: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 336: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 337: ......
Page 338: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...