CAPINTEC, INC.
CAPTUS
®
3000
Figure 5-15 Constancy Test Results Screen
An unacceptable deviation is indicated in red as HIGH.
If Deviation is indicated as HIGH, check the following:
•
The positioning of the rod source, especially for the Probe detector.
•
Ensure that there are no other sources (including dosed patients) in the area.
•
Click the
button or press
Alt+T
at the bottom of the Constancy Test screen.
Calibrated Activity and Calibration Date should match your Cs137 standard.
If a problem is found with any of the three items above, then correct the problem and
repeat the Constancy Test.
If no problems are found with the above items, then the efficiency of the Cs137 standard
should be re-measured. If this does not correct the problem, contact Capintec’s only
Authorized Service Center for service.
10. Click the
button or press
Alt+E
to exit and return to Figure 5-1 Quality Assurance
QUICK TIPS:
•
To print current results for record keeping, click the
button or press
Alt+P
on Constancy Test screen after the test is completed.
•
To view or print results for all the Quality Assurance tests performed to date,
click the
button or press
Alt+V
(Refer to the
section on page 5-24).
5-14
QUALITY ASSURANCE
July 14
Summary of Contents for CAPTUS 3000
Page 2: ......
Page 14: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 22: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 86: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 124: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 266: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 284: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 292: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 304: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 334: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 336: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 337: ......
Page 338: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...