CAPINTEC, INC.
CAPTUS
®
3000
of thyroid hormone, or Jod-Basedow disease. During the treatment of thyroid cancer by I-131
ablation, uptake measurements are used to monitor thyroid activity.
The CAPTUS
®
3000 system will store patient demographic and dosage information, count
data for both the administered dosage and the patient, decay correct all count data, calculate
thyroid uptake results, and store all of this information. Regions of Interest have been defined
for three isotopes, I-131, I-123, or Tc99m. If normal values have been entered in the setup,
then uptake results exceeding these values will be flagged. All of the patient information may
be printed. The thyroid uptake software has been designed to accommodate a variety of
different counting protocols, using either capsules or liquid dosages.
Note: In the following sections of the chapter, the word capsule will be used for either a
capsule or liquid dosage.
To access the Thyroid Uptake functions, click the
button or press
Alt+U
on
the CAPTUS
®
3000 Main screen. Figure 7-1 Main Thyroid Uptake Screen will appear. The
Patient Directory
box will be empty if there are no existing patients in the database. This
box will display the patient’s name, ID, administered isotope (if known) and the time and date
of dose administration (if already done). By default, patients are sorted by their dose
administration time, latest one first. The
Patient Status
box below the
Patient Directory
box
shows the current status of the selected patient. If the uptake procedure has been started,
the updated results will be displayed here.
Figure 7-1 Main Thyroid Uptake Screen
To return to the CAPTUS
®
3000 Main screen from Figure 7-1 Main Thyroid Uptake Screen,
click the
icon or press the
Esc
key.
7-2
THYROID UPTAKE
November 13
Summary of Contents for CAPTUS 3000
Page 2: ......
Page 14: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 22: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 86: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 124: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 266: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 284: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 292: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 304: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 334: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 336: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 337: ......
Page 338: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...