CAPINTEC, INC.
CAPTUS
®
3000
Figure 5-8
If you are performing the Constancy Test, Figure 5-9 will be displayed.
Figure 5-9
12. Place the Cs137 source rod in the selected detector. Since the Constancy Test is a
measure of reproducibility, it is
important
that the source be placed in exactly the same
position each time the Constancy Test is performed.
13. Click the
button to proceed with the test, or else click the
button
to stop the process. As the counting progresses, the live spectra and measurement times
are displayed.
14. To stop the counting, click the
button or press
Alt+A
. The collected data is
discarded. To re-start the process, click the
button or press
Alt+S
.
15. After the test is completed, the results of the test appear in the box below as shown in
Figure 5-10. Note the Deviation value at the bottom.
5-8
QUALITY ASSURANCE
July 14
Summary of Contents for CAPTUS 3000
Page 2: ......
Page 14: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 22: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 86: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 124: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 266: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 284: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 292: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 304: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 334: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 336: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 337: ......
Page 338: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...