A-5
Model 199 Differential Pressure Unit (DPU)
Section 1
Section 2—Theory of Operation
Basic Components
Low Pressure
Housing
HP Housing
Torque Tube
Shaft
Center Plate
Torque Tube
Low Pressure Bellows
Range Spring
High Pressure
Bellows
High Pressure
Overrange
Valve
Temperature
Compensator
Valve Stem
Low Pressure
Overrange
Valve
Housing
Bolts
Pulsation Damper
Damping Valve Plug
Figure 2.1—199 BUA cutaway
Pressure Housings
The two pressure housings of the Model 199 DPU are available in the various
safe working pressure ratings defined in
Table 1.1, page A-4,
and in the dimen-
sional drawings in Section 6.
Each pressure housing has two tapped connection ports: one port is located in
the top of the housing, the other port is located in the bottom of the housing.
Bellows
The bellows of the Model 199 DPU are available in the various materials
and sizes (refer to
Table 1.1, page A-4
) to accommodate the various calibration
ranges.