Sets the user defined waveform group that will be loaded at power on.
Available groups are 0, 1, 2 and 3. Each group can contain up to 50
user-defined waveforms. A waveform group can only be loaded at
power up. To change groups, you must change this field to the desired
new group and cycle the power to the AC source or issue a *RST
command over one bus.
CLOCK MODE
Sets the clock source used at power up. Available settings are Stand
Alone (STAND), MASTER (-LKM clock and lock master), and AUX
(-LKS clock and lock auxiliary).
NO. OUTPUT
This field is always set to THREE can cannot be changed. It reflects
the three phase output configuration of the BPS Sereis.
VOLT ALC
Determines ALC mode at power on. The ALC mode adjusts the
output voltage based on internal voltage measurement system and
provides enhanced output regulation and accuracy. Available settings
are ON, OFF or REG.