iL Series User Manual
68
Rev B
California Instruments
6.4.4 Calibration Configuration Switch
An internal set of DIP switches control access to calibration commands. The four
DIP switches are located on the A8 DSP board and are accessible by removing the
top cover. Switches 1 and 2 set the calibration configuration. Figure 20 shows the
three different switch settings for switches 1 and 2.
Switch Position
Description
Normal
This is the default switch position. The calibration functions
are accessible after entering a numeric password. The
default password is 0 and is changeable by the user.
Clear Password
The calibration password is reset to 0 and the remote
programming language is set to SCPI. This option is useful
if the user has forgotten the password.
Inhibit Cal
In this position the calibration of the power source cannot
be changed. All calibration commands are disabled. If the
CAL ON command is sent with the switch in this position,
an Out Of Range error will be displayed on the front panel.
This option is useful in installations where calibration
access is guarded by instrument seals.
Figure 20 Calibration Switches
Summary of Contents for 3000iL
Page 9: ...iL Series User Manual California Instruments Rev B ix page intentionally left blank...
Page 10: ......
Page 37: ...iL Series User Manual California Instruments Rev B 27...
Page 89: ...iL Series User Manual California Instruments Rev B 79 Figure 23 AC Power Source Block Diagram...
Page 93: ...iL Series User Manual California Instruments Rev B 83 Figure 25 AC Input Board Parts Location...
Page 94: ...iL Series User Manual 84 Rev B California Instruments Figure 26 Mother Board Parts Location...