iL Series User Manual
California Instruments
Rev B
63
The following parameters may be calibrated:
•
AC output voltage
•
realtime output voltage (4801iL only)
•
output voltage readback
•
over voltage protection
•
AC output current
•
output current readback (4801iL only)
•
output impedance (4801iL only)
6.4.1 Front Panel Calibration Menu
The Entry keypad is used for calibration functions.
Shift Calibration
Press this key to access the calibration menu list.
Display
Measurement
CAL ON <value>
turns calibration mode on when the correct password
value is entered
CAL OFF
turns calibration mode off
CAL:LEV <char>
advance to next step in sequence (P1, P2, P3 or P4)
CAL:DATA <value>
input a calibration measurement
CAL:VOLT:AC
begin AC voltage calibration sequence
CAL:VOLT:PROT
begin voltage protection calibration
CAL:CURR:AC
begin AC current programming calibration sequence
CAL:CURR:MEAS
begin current measurement calibration sequence,
4801iL only
CAL:IMP
begin output impedance calibration sequence, 4801iL
only
CAL:VOLT:RTIME
begin real-time voltage programming calibration, 4801iL
only
CAL:SAVE
saves the calibration constants in non-volatile memory
CAL:PASS <value>
set new calibration password
6.4.2 Front Panel Calibration
WARNING:
LETHAL VOLTAGES. AC sources can supply 425 V
peak at their output. DEATH on contact may result if the output
terminals or circuits connected to the output are touched when power
is applied.
Summary of Contents for 3000iL
Page 9: ...iL Series User Manual California Instruments Rev B ix page intentionally left blank...
Page 10: ......
Page 37: ...iL Series User Manual California Instruments Rev B 27...
Page 89: ...iL Series User Manual California Instruments Rev B 79 Figure 23 AC Power Source Block Diagram...
Page 93: ...iL Series User Manual California Instruments Rev B 83 Figure 25 AC Input Board Parts Location...
Page 94: ...iL Series User Manual 84 Rev B California Instruments Figure 26 Mother Board Parts Location...