iL Series User Manual
20
Rev B
California Instruments
These keys let you select the previous or the next parameter for a specific
command. If the command has a numeric range, these keys increment or
decrement the existing value.
←
Key
The backspace key is an erase key. If you make a mistake entering a digit and have
not yet pressed Enter, you can delete the digit by pressing
←
. Delete more digits by
repeatedly pressing this key.
Enter Key
Executes the entered value or parameter of the presently accessed command. Until
you press this key, the parameters you enter with the other keys are displayed but
not entered into the AC source. After pressing Enter, the AC source returns to Meter
mode in most cases. In Harmonic or List mode, the AC source displays the next
point in the list.
3.3 Output Checkout
WARNING:
LETHAL VOLTAGES. AC sources can supply 425 V
peak at their output. DEATH on contact may result if the output
terminals or circuits connected to the output are touched when power
is applied.
The output checkout test requires that you connect light bulbs to the output of the
unit and apply a potentially hazardous voltage of 120 Vac. Properly shield all
connections and wires.
The test in this section checks for output voltage and current on the AC source by
having you connect light bulbs to the output of the unit. The following equipment is
recommended for performing this output checkout procedure:
•
3 - 100 W /120 V light bulbs (use only one bulb for 4801iL)
•
3 - sockets
•
wires for connecting sockets to the unit
When the AC source is turned on as shipped from the factory, it asserts the *RST
state. You can subsequently program the unit to turn on according to the state stored
in *RCL location 0, as explained in section 4.11. The following procedures assume
that the unit turns on in the *RST state.
Make sure that the unit is turned off, and make the following connections to the
output.
Summary of Contents for 3000iL
Page 9: ...iL Series User Manual California Instruments Rev B ix page intentionally left blank...
Page 10: ......
Page 37: ...iL Series User Manual California Instruments Rev B 27...
Page 89: ...iL Series User Manual California Instruments Rev B 79 Figure 23 AC Power Source Block Diagram...
Page 93: ...iL Series User Manual California Instruments Rev B 83 Figure 25 AC Input Board Parts Location...
Page 94: ...iL Series User Manual 84 Rev B California Instruments Figure 26 Mother Board Parts Location...