iL Series User Manual
California Instruments
Rev B
61
Step Action
Expected
Result
7.
Change the voltage range of the unit from the 300 volt
range to the 150 volt range.
8.
Program the output for 5 volts and the High Current value
specified in Table 41.
9.
Observe the DVM voltage reading. Divide this by the
resistance of the current monitor resistor. Record the
result as the High Current value in Table 41.
Value within
high current
limits
specified in
Table 41.
10.
Record the front panel display readback.
Value within
readback
limits
specified in
Table 41
11.
Disable the output (Output On/Off or OUTP OFF), turn off
the power, and remove the shunt from across the load.
12
If the AC source you are verifying is in three-phase mode,
repeat steps 1 through 11 for output phases 2 and 3.
Press Phase Select to select the next phase.
Table 40 : Current programming readback test steps