PN MN236 A | User Manual    7 - ENG

Sprint / Stroller / Hi Flow Stroller

2. Make sure the flow control knobs on both the 

Sprint/Stroller/Hi Flow Stroller and reservoir are 

in the off (0) position

3. Make the Sprint/Stroller/Hi Flow Stroller fill con-

nector is in the upright position to the stationary 

connector.

4.  Rotate the Sprint/Stroller/Hi Flow Stroller coun-

terclockwise (   ) until you feel the pin and slot 

engage.

5. Carefully and firmly rotate the Sprint/Stroller/Hi 

Flow Stroller back to the upright position. Now 

the units are locked together. 

NOTE: Continued on step 6.

NOTE: For push-on style connector, hold the unit in 
place throughout the fill.

6. Open the vent valve to begin filling your Sprint/

Stroller/Hi Flow Stroller. You should hear some 

oxygen escaping; do not worry. The hissing sound 

you hear is perfectly normal.

NOTE: Liquid level gauge readings are most accurate 
30 seconds or more after the Sprint/Stroller/Hi Flow 
Stroller is filled.

7. You may also see vapor around the connection. 

This is normal and is due to the great difference 

in temperature between the liquid oxygen and the 

warm air surrounding the units.

8. When the unit is full, close the vent valve. You 

can tell that the unit is full when the hissing sound 

changes in tone, and some liquid comes out of the 

vent line.

NOTE: Closing the valve after a nearly continuous 
stream of liquid is discharging fills the Sprint/Stroller/
Hi Flow Stroller most completely, but wastes some 
liquid oxygen.
NOTE: Closing the valve at the first sign of liquid 
discharge does not fill the Sprint/Stroller/Hi Flow 
Stroller as completely, but saves liquid oxygen.

9. For the push-on style connectors, push the re-

lease button down until the units separate. For all 

other connectors rotate the unit until it separates 

from the stationary unit.

Portable Release Button (push-on style only)

Summary of Contents for SPRINT

Page 1: ...Sprint Stroller Hi Flow Stroller User Manual English...

Page 2: ...Warning Reg W001 Council Directive 93 42 EEC concerning medical devices Authorized representative in the European Community This device complies with the requirements of Directive 93 42 EEC concerning...

Page 3: ...pacity 513 L 1026 L 1025 L Weight Empty 4 5 lb 2 04 kg 5 0 lb 2 27 kg 5 5 lb 2 49 kg Weight Filled 6 0 lb 2 72 kg 8 0 lb 3 63 kg 8 5 lb 3 86 kg Height 11 75 in 298 mm 13 5 in 343 mm 13 5 in 343 mm Dim...

Page 4: ...on the order of a physician The unit contains liquid oxygen which is extremely cold almost 300 F 184 C Exposure to such a low temperature can cause severe frostbite Liquid and gaseous oxygen though no...

Page 5: ...capabilities It is recommended to have an alternate source of supplemental oxygen in the event of mechani cal failure These portable devices also have a quick disconnect valve that can be coupled with...

Page 6: ...off 0 position 3 Position the fill connector in the upright position over the stationary unit s connector 4 a If your connector is a push on Lower the Sprint Stroller Hi Flow Stroller unit on to the...

Page 7: ...al NOTE Liquid level gauge readings are most accurate 30 seconds or more after the Sprint Stroller Hi Flow Stroller is filled 7 You may also see vapor around the connection This is normal and is due t...

Page 8: ...the Sprint Stroller Hi Flow Stroller is being refilled immediately after a period of use fill it and then wait 10 minutes after filling with the FCV set at zero for the pressure to stabilize Then oper...

Page 9: ...oxygen in the unit with the electronic liquid level gauge Gen 3 Gauge Depress the push button liquid level switch on top of the unit for two seconds minimum Read across the top of the light bar to in...

Page 10: ...sition Now replace the cup as previ ously described For soft case Unzip the bottom zipper and pull back the bottom cover Remove the pad from the cup and wring out any absorbed moisture Allow the pad t...

Page 11: ...e off 0 position The Portable does not fill Verify that there is oxygen in the Reservoir Ensure that the Portable and Reservoir fill connectors are fully engaged through out the filling process The Po...

Page 12: ...l Always return Sprint Stroller Hi Flow Stroller in cluding all components to your homecare provider for proper disposal You can also contact your local city or town offices for instructions on proper...

Page 13: ...gnetic compatibility EMC and needs to be installed and put into service according to the EMC information provided in this manual Caution Portable and mobile radio frequency RF communications equipment...

Page 14: ...standard requirement for equipment which has been tested to specific test levels and over specific frequency ranges and been found compliant with regulations Table 3 Guidance and manufacturers declara...

Page 15: ...om fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equip ment marked...

Page 16: ...h no further obligation or consequence All rights not expressly stated herein are reserved by us as applicable www cairemedical com CAIRE Inc 2200 Airport Industrial Dr Ste 500 Ball Ground GA 30107 U...

Reviews: